AIMSIZER SCIENTIFIC PTE. LTD.

201. Measurement difference between a fisher sub sieve sizer(FSSS) and a laser particle size analyzer(LPSA)?

201. Measurement difference between a fisher sub sieve sizer(FSSS) and a laser particle size analyzer(LPSA)?

-The results obtained by a fisher sub sieve sizer is average particle size DSV, which is similar to test results D(3,2) obtained by a laser particle size analyzer. However a FSSS actually measure but a LPSA gives particle size distribution before calculating the D(3,2) assuming that the particles are
spherical. That is to say spherical particles give smaller difference and irregular particles give bigger difference.

Fisher Scientific Sub-sieve Sizer Parts-201pic

HMK-22 Fisher sub sieve sizer

Fisher sub sieve sizer https://www.aimsizer.com