Air Jet Sieve Analyses Lactose Pharmatose 200M Particle Size
Air Jet Sieve Analyses Lactose Pharmatose 200M Particle Size
Recommended Equipment
- Hosokawa Alpine Air Jet Sieve
- Air Jet sieves as tabulated below
- Balance with readability of 0.1g
Sieve table
| Lactose | Sieve required | 
| Pharmatose 50M | 200um,400um | 
| Pharmatose 80M | 100um,250um,315um | 
| Pharmatose 90M | 63um,100um,150um,315um | 
| Pharmatose 100M | 63um,150um,250um | 
| Pharmatose 110M | 63um,100um,150um,315um | 
| Pharmatose 125M | 45um,63um,100um,150um | 
| Pharmatose 150M | 45um,100um,150um,315um | 
| Pharmatose 200M | 45um,100um,150um,250um | 
| Pharmatose 350M | 45um,100um,250um | 
| Pharmatose 450M | 45um,63um,150um | 
Procedure
- Weigh 20.0 gram of the sample to be tested onto the required sieve (weight b). For milled products (150M,200M,350M and 450M) it is advisable to add about 0.1 g of aluminium oxide to the sample on the sieve to avoid lumping.
- Sieve the sample for 3 minutes using the Hosokawa Alpine Air Jet Sieve operating at 200 mm of water underpressure
- Determine the weight of the residue (weight a)
For each sieve, calculate the percentage undersize as: % underside =100*(b-a)/b
Air jet sieve https://www.aimsizer.com